Opened 8 years ago
Last modified 6 years ago
#850 closed defect
Custom pinhole smearing is broken? — at Initial Version
Reported by: | ajj | Owned by: | |
---|---|---|---|
Priority: | blocker | Milestone: | SasView 4.1.0 |
Component: | SasView | Keywords: | |
Cc: | Work Package: | SasView Bug Fixing |
Description
Setting values in custom pinhole smearing gives a different answer to loading the same values from a data file.
Some instruments do not provide resolution information - hence the need for the custom smearing.
Taking data set and calculating a, say, 8% dQ/Q and putting that in the file as the 4th column behaves as expected.
Generating a custom pinhole smear with *the same dQ_low and dQ_high* as the dataset with "fake" resolution column gives a completely different answer and blows the curve out.
Attached images are for the attached data set with
(a) resolution from dQ data which was faked in the file.
(b) resolution from a custom pinhole smear using the same dQ_low and dQ_high
Data file with faked dQ values at 8% dQ/Q